• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/31/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that after a lightning storm patient's transmitter did not power up.Upon removing the prongs, patient noted that one of the green contacts was burnt.A new transmitter was ordered.
 
Manufacturer Narrative
Final analysis found burn marks on the main printed circuit board and inner casing of ac power adapter.It was concluded that the damage sustained occurred after exposure to a high current flow from the power outlet caused by the lightning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7783508
MDR Text Key117211133
Report Number2017865-2018-11822
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-