Catalog Number M003IC0581150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.Analysis of the returned device revealed that the distal access catheter (dac) shaft was broken at the junction of the grilamid sleeve and the catheter shaft adjacent to the hub.Functional testing could not be performed due to the condition of the returned device.It is probable that the device was damaged as a result of handling of the device during removal of the device from its packaging, therefore an assignable cause of undeterminable will be assigned to this investigation.
 
Event or Problem Description
During the analysis of the returned device, it was revealed that distal access catheter (dac) shaft was broken.There was no clinical consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS CATALYST 5 - 058X115 (FDA)
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7783598
Report Number3008881809-2018-00344
Device Sequence Number1258399
Product Code DQY
UDI-Device Identifier07613327012019
UDI-Public07613327012019
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K151667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue NumberM003IC0581150
Device Lot Number19449133
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2018
Initial Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-