| Model Number |
M004PM45000 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the returned device matches with upn and lot provided by the customer.Some residues of char were observed on the edge of the tip in distal end.The steering knob and the tension control knob functioned properly on both lock and unlock positions.The electrical test was performed in neutral configuration and the device was found within specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is anticipated procedural complication as the event is due to a known physiological effect of the procedure noted within the directions for use, and/or device labeling.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a steam pop occurred.A intellatip mifi xp was used in a supraventricular tachycardia ablation.On the third ablation delivery the physician experienced a steam pop.Procedure was completed using this device.No patient complications were reported.
|
| |
|