| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 13-aug-2018 with the following findings: during testing, the pump was powered on and emitted a cs 069 call service alarm immediately upon attempting a rewind step.The call service alarm was recorded in the black box data as a cs 087 failure.Investigation revealed that a language corruption occurred at a component on the printed circuit board resulting in a call service alarm.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|