• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Codes High Test Results (2457); Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# ((b)(4).Note: manufacturer contacted customer on (b)(4) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.Customer is concerned with tests results from meter to meter comparison of 42 mg/dl using truemetrix meter and 80 mg/dl using another device; customer is also concerned with results obtained of 65, 84, 51 and 69 mg/dl.The customer's expected fasting blood glucose test result range is 80 - 120 mg/dl.Customer stated that the result of 42 mg/dl was not in the meter memory as the meter is also not storing the results.Customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 03/31/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).Customer says that result is not in meter memory because meter is also not storing results.Customer says he ran controls before calling and results were within range but is still unsatisfied with the meter.Customer declined a replacement meter.Customer called back on (b)(65) 2018 and reported complaint for high and low blood glucose results.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.The test strip lot manufacturer's expiration date is 03/31/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Customer called a while back regarding low readings.After performed troubleshooting, customer decided to continue using true metrix meter.While doing follow up today, customer says that he is now concerned that meter is just reading out of range and is inaccurate.Reviewed meter with customer.Customer had a hard time with times and dates.Not sure if they are correct.Customer wants meter replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7784619
Report Number1000113657-2018-00863
Device Sequence Number16656685
Product Code NBW
UDI-Device Identifier00021292006068
UDI-Public(01)00021292006068
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/31/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2635
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2018
Initial Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY
-
-