• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; LIGACLIP APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; LIGACLIP APPLIER Back to Search Results
Catalog Number LX107
Device Problem Device Slipped (1584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Only event year known: 2018.Device analysis: the analysis results confirmed that the lx207 device was returned nonfunctional as the jaws were misaligned.Functional test was performed on one clip, it was loaded and retained as intended, but due to the damaged jaws, the clip formation was not proper.No conclusion could be reached as to what may have caused the found condition of the jaws.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use." device was received as a lx207 instead of the reported lx107 code.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment.  the certificate records are accessible through external manufacturing. .
 
Event Description
It was reported that the clip applier was not holding clips.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
LIGACLIP APPLIER
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7784628
MDR Text Key117282290
Report Number3005075853-2018-12113
Device Sequence Number1
Product Code GDO
UDI-Device Identifier20705036012853
UDI-Public20705036012853
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLX107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Initial Date Manufacturer Received 08/03/2018
Initial Date FDA Received08/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/1988
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-