| Model Number |
PBX-SYS-AFFIRM-3D |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As of today the investigation is still in process.A follow up will be submitted as needed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy exam, the c-arm released while the needle was in the patient's breast.No injury reported.The patient did not require additional patient care than the normal patient care post vacuum biopsy.
|
| |
|
Additional Manufacturer Narrative
|
|
As of today the investigation is still in-process.We are awaiting the final resolution and part return for evaluation.A supplement will be filed as needed.
|
| |
|
Event or Problem Description
|
|
The field engineer worked with technical support and it was determined that the detent bracket, dc motor, and spring needed to be replaced.Once this was completed the system was working as intended.
|
| |
|