• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVUITY, INC SINGLE FIBER OPTIC CABLE, 10FT, PINK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVUITY, INC SINGLE FIBER OPTIC CABLE, 10FT, PINK Back to Search Results
Model Number FC1SP
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During evaluation, it was noted that broken fibers were present throughout the cable.Additionally, the cable exhibited a concentrated number of broken fibers which indicates mishandling of the cable.The concentration of broken fibers led to an elevated temperature.Invuity's ifu clearly states to handle the cables with care and not to coil the cables tighter than six (6) inches.
 
Event or Problem Description
It was reported during a hip replacement procedure, the fiber optic cable may have been placed on the drape of the patient.During use, the temperature of the fiber optic cord felt hot and left a burn mark on the drape.The case was completed without incident or product malfunction.There was no patient injury involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE FIBER OPTIC CABLE, 10FT, PINK
Common Device Name
SINGLE FIBER OPTIC CABLE, 10FT, PINK
Manufacturer (Section D)
INVUITY, INC
444 de haro st
san francisco CA 94107
MDR Report Key7784680
Report Number3011394215-2018-00006
Device Sequence Number2217903
Product Code FST
UDI-Device Identifier00816728020202
UDI-Public00816728020202
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberFC1SP
Device Catalogue NumberFC1SP
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2018
Initial Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-