BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number ATG80224
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had difficulty inflating and deflating.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was returned for evaluation.The device was attached to an in-house inflation device and inflated without issue.The device was then able to be fully deflated within the specified time constraints.Therefore, the investigation is unconfirmed for the reported inflation and deflation issues.The definitive root cause for the reported issues could not be determined based on the available information.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(b)(4).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had difficulty inflating and deflating.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7784996
Report Number2020394-2018-01461
Device Sequence Number19387650
Product Code LIT
UDI-Device Identifier00801741060977
UDI-Public(01)00801741060977
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K122984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATG80224
Device Lot Number93JZ0173
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date08/15/2018
Supplement Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No