• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON KEELED GLENOID, SZ 50MM, EPLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON KEELED GLENOID, SZ 50MM, EPLUS Back to Search Results
Model Number 521-01-150
Health Effect - Clinical Codes Pain (1994); Rupture (2208)
Date of Event 07/25/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Third revision surgery - due to the patient having pain and a subscapular rupture.The pain plan was to remove the neck, head, and glenoid and revise to a reverse total shoulder.
 
Additional Manufacturer Narrative
(b)(4).Manufacturer narrative: the reason for this revision surgery was due to instability and a subscapular rupture.The previous revision surgery and the revision surgery detailed in this investigation occurred 7.8 months apart.The in vivo time for the implanted device is 1.5 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to instability and subscapular rupture.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Third revision surgery - due to the patient having instability and a subscapular rupture.The pain plan was to remove the neck, head, and glenoid and revise to a reverse total shoulder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON KEELED GLENOID, SZ 50MM, EPLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7786631
Report Number1644408-2018-00742
Device Sequence Number1287820
Product Code PAO
UDI-Device Identifier00888912145343
UDI-Public(01)00888912145343
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Model Number521-01-150
Device Catalogue Number521-01-150
Device Lot Number796N1020
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer09/07/2018
Initial Report FDA Received Date08/15/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
-
-