| Catalog Number |
210110 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Sled would not tighten enough to keep leg upright.Leg keep falling either internal or external.Surgical delay yes -16-30 minutes; case type: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported issue it was reported that the ball is loose.Product history review of the device history records indicate (b)(4) devices were manufactured and accepted into final stock on 05/06/2017.No non-conformances were identified during inspection.Complaint history review: a review of complaints related to p/n 210110, lot 19620816 shows no additional complaint(s) related to the failure in this investigation.Material analysis: material analysis was not completed as no material failure was alleged.Functional inspection: the sled assembly was tested using a known good base bar and boot assembly.The issue could not be reproduced after the lever was locked conclusion: the issue could not be reproduced.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there has been no nc or capa associated with the product and failure mode reported in this event.
|
| |
|
Event or Problem Description
|
|
Sled would not tighten enough to keep leg upright.Leg keep falling either internal or external.Surgical delay yes -16-30 minutes.Case type: tka.
|
| |
|
Search Alerts/Recalls
|