• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. HIGHLY CROSS LINKED KNEE; DOMED TRI-PEG PATELLA, 32X8MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. HIGHLY CROSS LINKED KNEE; DOMED TRI-PEG PATELLA, 32X8MM, E-PLUS Back to Search Results
Model Number 130-03-732
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Date 07/26/2018
Event Type  Injury  
Event Description
Revision surgery - due to the implant possibly being loose, but not sure.All djo implants came out of the patient and a competitor brand implants went in.
 
Manufacturer Narrative
The reason for this revision surgery was due to possible loose implant.The previous surgery and the surgery detailed in this investigation occurred 1.6 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to possible loose implant.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Manufacturer Narrative
Additional narrative: concomitant medical products.Additional concomitant items needed to be added to the mdr complaint.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGHLY CROSS LINKED KNEE
Type of Device
DOMED TRI-PEG PATELLA, 32X8MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7786710
MDR Text Key117254004
Report Number1644408-2018-00740
Device Sequence Number1
Product Code OIY
UDI-Device Identifier00888912108539
UDI-Public(01)00888912108539
Combination Product (y/n)N
PMA/PMN Number
K113756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/02/2021
Device Model Number130-03-732
Device Catalogue Number130-03-732
Device Lot Number155N1199
Was Device Available for Evaluation? No
Date Manufacturer Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
242-02-107, LOT 870T1006; 343-10-706, LOT 010U1004; 351-02-106, LOT 992R1038; 402438, LOT 871270
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-