(b)(4).Information regarding patient weight, race, ethnicity, and medical history were not provided.Procode is krd/hcg.(b)(4).The name, phone and email address of the initial reporter is unknown.[conclusion]: the healthcare professional reported that during the procedure to treat an unruptured aneurysm in a (b)(6) female patient, during the delivery of the deltafill 18 3 mm x 12 cm coil (dlf180312 / l10330), the coil prematurely detached before it was fully inserted into the aneurysm.The physician was able to push the coil into the aneurysm using the coil delivery wire and the coil was implanted.The event did not result in any clinical delay.There was no report of patient injury or adverse event associated with the reported issue.Based on complaint information the deltafill 18 3 mm x 12 cm coil remains implanted and is thus, not available to be returned for analysis.A review of manufacturing documentation associated with this lot (l10330) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no non-conformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Since the product was not returned for analysis, no product failure analysis can be conducted, and no determination of causes and possible contributing factors could be made.As such the investigation will be closed.With the limited information available and without the product being available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and/or device manipulation/interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during the procedure to treat an unruptured aneurysm in a (b)(6) female patient, during the delivery of the deltafill 18 3 mm x 12 cm coil (dlf180312 / l10330), the coil prematurely detached before it was fully inserted into the aneurysm.The physician was able to push the coil into the aneurysm using the coil delivery wire and the coil was implanted.The event did not result in any clinical delay.There was no report of patient injury or adverse event associated with the reported issue.
|