MEDIVATORS HYBRID IRRIGATION TUBING; ENDOGATOR TUBING Back to Search Results
Model Number 100609
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aspiration/Inhalation (1725)
Date of Event 07/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that a patient aspirated during an upper gi endoscopy procedure using an olympus hq190 endoscope and was hospitalized for aspiration pneumonia.Medivators endogator hybrid irrigation tubing was being used during the procedure and the facility reported that the tubing was leaking and caused the patient adverse event.The facility reported their endogator hybrid irrigation tubing was leaking during an upper gastrointestinal procedure and caused the patient to aspirate.The endogator hybrid tubing does not come into contact with the patient.The hybrid tubing connects to the sterile water bottle and connects to the scope to providethe flow of water and co2 through the endoscope.The air/water tubing has a backflow valve.It is not anatomically possible for the tubing to be the cause of the aspiration.Medivators field staff and regulatory has visited this facility to observe their set up and use of the product.Medivators personnel confirmed that the facility is using the tubing in accordance with the instructions for use and did not witness the product leaking or any other product related issues.It was observed that there is inconsistency of set up in one procedure room to the next with regards to irrigation pump brand and water bottle brands being used to know what could potentially cause a leak.It has been requested product returned for internal quality investigation.Medivators field staff remain in close contact with this facility and will continue to monitor this account.This has been documented and is being monitored in medivators complaint system.
 
Event or Problem Description
It was reported that a patient aspirated during an upper gi endoscopy procedure using an olympus hq190 endoscope and was hospitalized for aspiration pneumonia.Medivators endogator hybrid irrigation tubing was being used during the procedure and the facility reported that the tubing was leaking and caused the patient adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYBRID IRRIGATION TUBING
Common Device Name
ENDOGATOR TUBING
Manufacturer (Section D)
MEDIVATORS
3150 pollok drive
conroe TX 77303
MDR Report Key7787005
Report Number2150060-2018-00052
Device Sequence Number19284802
Product Code OCX
UDI-Device Identifier00677964052354
UDI-Public00677964052354
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K033695
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other,Patient Monitoring
Type of Report Initial
Report Date (Section B) 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number100609
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2018
Initial Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
-
-