• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035421020
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.During the analysis of the returned device, it was revealed that the main coil was broken, stretched and the main coil suture was damaged.In addition, the coil delivery wire was kinked due to handling.The functioning test could not be performed due to the condition of the returned device.Additional information indicated that the coil was confirmed to be in good condition prior to use and it was prepared as per dfu (direction for use).It is probable that the device was damaged during the clinical procedure due to some procedural/anatomical factors encountered during use.Therefore, an assignable cause of procedural factors will be assigned to the analyzed defects on the main coil and the suture.The issue is associated with a product that meets the design and manufacture specifications and was used according to the dfu but due to procedural and/or anatomical factors during use, the device performance was limited.
 
Event or Problem Description
During the analysis of the returned device, it was revealed that the main coil was broken.No clinical consequences reported to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7787175
Report Number3008881809-2018-00349
Device Sequence Number181832
Product Code HCG
UDI-Device Identifier07613252600329
UDI-Public07613252600329
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue NumberM0035421020
Device Lot Number20267740
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2018
Initial Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-