• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-5312
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 07/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device has not been returned yet.
 
Event or Problem Description
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
 
Event or Problem Description
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
 
Additional Manufacturer Narrative
This is a duplicate report.This complaint was previously reported under report 3006630150-2018-02328.
 
Event or Problem Description
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
MDR Report Key7787210
Report Number3006630150-2018-60606
Device Sequence Number227884
Product Code LGW
UDI-Device Identifier08714729764465
UDI-Public08714729764465
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSC-5312
Device Catalogue NumberSC-5312
Device Lot Number17152580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer09/03/2018
10/05/2018
Initial Report FDA Received Date08/15/2018
Supplement Report FDA Received Date09/28/2018
10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-