| Model Number |
SC-5312 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device has not been returned yet.
|
| |
|
Event or Problem Description
|
|
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
|
| |
|
Event or Problem Description
|
|
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a duplicate report.This complaint was previously reported under report 3006630150-2018-02328.
|
| |
|
Event or Problem Description
|
|
A report was received that the patient ipg pocket site felt hot while charging.Physician assessed that the patient had a low temperature burn that resulted in a blister.A gauze dressing was applied and patient was advised to discontinue use of the device for one week.
|
| |
|
Search Alerts/Recalls
|