| Model Number |
5586 |
| Medical Device Problem Code |
Pacemaker Found in Back-Up Mode (1440)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that when an asymptomatic patient presented in clinic during follow-up, device was found to be in backup vvi mode.It was noted that the patient had been lost to follow-up since post-implant device check.Device was explanted and replaced due to normal eri.Patient¿s condition was good post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Field complaint of backup operation was confirmed and was due to normal battery depletion.Device was tested on bench and no anomalies were noted.
|
| |
|