• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER Back to Search Results
Catalog Number CRU14SA
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review:the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: one crosser catheter was returned for evaluation.A complete core wire fracture and tip detachment were noted at the distal end of the catheter.Therefore, the investigation can be confirmed for the alleged tip detachment.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review:the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Expiration date (04/2020).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that after approximately one minute and thirty-six seconds of activation in the sfa, the tip of the recanalization catheter allegedly detached.Reportedly, a straight flush catheter and guidewire were used to retrieved the tip.It was further reported that wire and catheter were used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7787595
Report Number2020394-2018-01477
Device Sequence Number228082
Product Code PDU
UDI-Device Identifier10801741125567
UDI-Public(01)10801741125567
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCRU14SA
Device Lot NumberGFCR1457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2018
Initial Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
-
-