MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 07/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implant and explant dates: if implanted or explanted, give date: not applicable, as the lens was inserted and removed.(b)(4).Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient's capsular bag torn while inserting a zcb00 16.5 diopter lens into the patient's operative eye.There was an incision enlargement made, a vitrectomy performed, and a suture was used.The replacement lens was model z9002 with an unknown diopter power.Reportedly, the patient is doing well.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7787657
Report Number3011852734-2018-00145
Device Sequence Number16698680
Product Code HQL
UDI-Device Identifier05050474531642
UDI-Public(01)05050474531642(17)220208
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/08/2022
Device Model NumberZCB00
Device Catalogue NumberZCB0000165
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date08/15/2018
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-