• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TENODESIS SCREW, PEEK; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TENODESIS SCREW, PEEK; SCREW, FIXATION, BONE Back to Search Results
Model Number TENODESIS SCREW, PEEK
Health Effect - Clinical Code No Information (3190)
Date of Event 07/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The device was reported to have been discarded by the facility.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a wrist joint ligament reconstruction procedure, the tip of inserter for the peek tenodesis screw broke off while inserting the screw.The broken tip was lodged i the head of the screw and was not able to be retrieved and is still in the patient.*no further details.
 
Additional Manufacturer Narrative
Complaint confirmed, evaluation of the device shows that 0.88in of the device extremity is missing.The most likely cause(s) of this type of event include flexing the joint during insertion of the implant with the driver engaged or prying/leveraging the driver while the implant is still loaded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENODESIS SCREW, PEEK
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7788236
Report Number1220246-2018-00598
Device Sequence Number14065197
Product Code HWC
UDI-Device Identifier00888867018662
UDI-Public00888867018662
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberTENODESIS SCREW, PEEK
Device Catalogue NumberAR-1530PS
Device Lot Number10168970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2019
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer07/25/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date01/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-