|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint devices used in the procedure were not returned for evaluation to mitek complaints and therefore we cannot confirm this complaint.However, three unused, unopened devices from customer¿s inventory were received and evaluated.Visual observation under magnification reveals no structural anomalies on the anchor.The assembly of the anchor is in accordance with the specs.Two out of the three anchors were deployed into a sample test foam successfully.As reported in the case, the anchors were tapped into the test hole and presented no structural damage.It was reported that the anchor placement was not off axis.From the information available, we cannot discern a root cause for this failure.A lot review was conducted and our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
The sales rep reported that during a shoulder procedure two of the customer's bioknotless rapide br w/orthocord broke during insertion.The sales rep reported that the surgeon did not use any unusual angles, that he drilled the bone hole first, and then used the hole finder next.The surgeon then used a mallet to insert the anchor 1/2 way into the bone hole then tapped the anchor in to finish inserting the anchor the rest of the way in when it broke into pieces.The surgeon repeated the procedure with another like device to the same exact effect, breaking into pieces once half way inserted.Both anchors were fully removed.Both anchors were of the same lot number.The surgeon completed the procedure with a third like device but different lot number, with no patient consequences.There was a 10 minute delay in the procedure.The sales rep stated that the surgeon had used mitek products a long time.The sales rep removed three remaining unopened, unused anchors from the customer that were of the same lot number and is returning these for evaluation.Please see also medwatch report 1221934-2018-50970.
|