Model Number STAY SAFE MTS SET
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A peritoneal dialysis (pd) patient reported that the stay safe mts set tubing seems to leak during peritoneal dialysis treatment.There was no indication of the patient encountering cycler alarms.Follow up with the patient¿s peritoneal dialysis nurse indicates that the patient did not experience any complications as a result of the alleged event.Additional information was solicited.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
".".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAY SAFE MTS SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
MDR Report Key7788632
Report Number8030665-2018-01241
Device Sequence Number1287205
Product Code KDI
UDI-Device Identifier00840861100699
UDI-Public00840861100699
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K904806
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSTAY SAFE MTS SET
Device Catalogue Number050-50758
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Received by Manufacturer 07/31/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date08/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; LIBERTY CYCLER; LIBERTY CYCLER SET 
-
-