Model Number TTA2
Medical Device Problem Code Patient Data Problem (3197)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although this issue occurs when a study is not closed prior to beginning the next study and has since been resolved in a newer lot, a thorough evaluation of this event will be included in a follow-up report upon investigation completion.
 
Event or Problem Description
A customer reported a patient data error while using the tomtec-arena.The software allowed patient data acquired from one data set to be included in the next patient study.There was no misdiagnosis or injury associated with this event.
 
Additional Manufacturer Narrative
This action was reported to fda per 21 cfr part 806 on 10/2/2019.Reference corrections and removal report number 8043836-10/02/19-001-c.
 
Additional Manufacturer Narrative
This action was reported to fda per 21 cfr part 806 on 10/2/2019.Reference corrections and removal report number (b)(4).
 
Additional Manufacturer Narrative
A thorough technical investigation was performed which included evaluation of workflow details, a software design issue was determined to be to cause and a solution has been included in a newly released software revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOMTEC-ARENA TTA2
Common Device Name
IMAGE REVIEW AND QUANTIFICATION SOFTWARE
Manufacturer (Section D)
TOMTEC IMAGING SYSTEMS GMBH
edisonstrasse 6
unterschleissheim, 85716
GM  85716
MDR Report Key7788665
Report Number8043836-2018-00001
Device Sequence Number1188371
Product Code LLZ
UDI-Device Identifier04260361440143
UDI-Public(01)04260361440143(11)160809(10)20.10
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K150122
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Administrator/Supervisor
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTTA2
Device Lot Number20.10
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer07/19/2018
07/19/2018
07/19/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date11/21/2019
12/20/2019
05/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
8043836-10/02/19-001-C
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-