STRYKER ORTHOPAEDICS-MAHWAH SECUR-FIT MAX 127 HIP STEM #8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROU Back to Search Results
Catalog Number 6052-0830S
Medical Device Problem Codes Degraded (1153); Device Damaged by Another Device (2915); Material Deformation (2976); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 07/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the reported lot.There have been no other similar events for the sterile lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned / not received.
 
Event or Problem Description
It was reported that patient's hip was revised due to impingement.Surgeon also reported suspicion of infection with evidence of poly wear cited as the cause.Intra-operatively, no visual evidence of poly wear was found.Patient was revised to a spacer including a 6 x 127 hfx stem.The trunnion of the stem was impinging with the cup.Sales rep believes that infection was confirmed to prompt implantation of a first stage antibiotic spacer.
 
Event or Problem Description
It was reported that patient's hip was revised due to impingement.Surgeon also reported suspicion of infection with evidence of poly wear cited as the cause.Intra-operatively, no visual evidence of poly wear was found.Patient was revised to a spacer including a 6 x 127 hfx stem.The trunnion of the stem was impinging with the cup.Sales rep believes that infection was confirmed to prompt implantation of a first stage antibiotic spacer.
 
Additional Manufacturer Narrative
An event regarding rom (impingement) and infection was reported.The rom (impingement) was confirmed through photographs provided.The infection was not confirmed method & results: device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted securfit stem.From the photographs provided there is evidence of deep gouging on the neck of the device.Medical records received and evaluation: no medical records were received for review with a clinical consultant device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot or sterile lot.Conclusion: the exact cause of the event (infection) could not be determined because insufficient information was provided.While the rom (impingement) event was confirmed through photographs provided the root cause could not be determined.Additional information, including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECUR-FIT MAX 127 HIP STEM #8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7789288
Report Number0002249697-2018-02559
Device Sequence Number1175402
Product Code LZO
UDI-Device Identifier07613327021110
UDI-Public07613327021110
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K051738
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2012
Device Catalogue Number6052-0830S
Device Lot NumberDY3MPD
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR
Patient Weight84
-
-