An event regarding crack/fracture involving a triathlon insert was reported.The event was confirmed through return of the fractured device.Method & results: device evaluation and results: material analysis carried out on the returned device concluded: burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.The posterior end of the medial condyle had fractured in overload.Explantation damage was also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: review of the provided medical records by a clinician indicated: no clinical or past medical history, no operative reports, and no serial x-rays or lateral views are available for review.No determination of range of motion or stability of the right knee prior to the revision is noted.In this morbidly obese male patient, after more than six years in situ, failure of the posterior aspects of the tibial insert in fatigue suggests soft tissue imbalance resulting in overload failure.No material or manufacturing discrepancies were noted in the material analysis report.Device history review: indicates all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: material analysis carried out on the returned device concluded: burnishing, scratching and third-body indentations were observed on the insert.These are common damage modes of uhmwpe.The posterior end of the medial condyle had fractured in overload.Explantation damage was also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.No clinical or past medical history, no operative reports, and no serial x-rays or lateral views are available for review.No determination of range of motion or stability of the right knee prior to the revision is noted.In this morbidly obese male patient, after more than six years in situ, failure of the posterior aspects of the tibial insert in fatigue suggests soft tissue imbalance resulting in overload failure.No material or manufacturing discrepancies were noted in the material analysis report.While the event was confirmed through the return of the device and material analysis could confirm no identifiable materials or manufacturing discrepancies, without clinical or past medical history, operative reports, and serial x-rays or lateral views the clinician could not confirm a root cause for this event.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|