MAUDE Adverse Event Report: REFER TO SECTION H10 MICRUSFRAME10 3MM X 5.4CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REFER TO SECTION H10 MICRUSFRAME10 3MM X 5.4CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number MFR100305
Medical Device Problem Codes Entrapment of Device (1212); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Information regarding patient age, gender, weight, medical history, race, and ethnicity was not provided.Date of event: the event occurred in 2018; however, the month and day were not reported.Procode: krd/hcg.Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)) the product is not available for evaluation and testing.Additionally, the sterile lot number is not known; therefore, a review of the manufacturing documentation could not be performed.Review of the imaging provided by the customer is pending at this time.Additional information will be submitted within 30 days of receipt.(b)(6).Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)) missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by a healthcare professional, during a coil embolization of a 3.5 x 2.8 x 2.7mm ruptured aneurysm, the 3mm x 5.4cm micrusframe10 (mfr100305/unk lot) thermo-mechanical framing coil appeared to fit perfectly in the beginning of the case, but the proximal portion of the embolic coil failed to break.The coil was supplied with a 3mm non-cerenovus coil, which was considered too small.It was further stated that "the last part of the coil did not perform loops anymore".There was no report of consequence of impact to the patient.Imaging was provided by the customer.The product will not be returned for analysis.It was reported that the information provided for this complaint was all of the details that were known/available regarding this event.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Corrected sections: h6 - device code - remove 1212.Complaint conclusion: as reported by a healthcare professional, during a coil embolization of a 3.5 x 2.8 x 2.7mm ruptured aneurysm, the 3mm x 5.4cm micrusframe10 (mfr100305/unk lot) thermo-mechanical coil appeared to fit perfectly in the beginning of the case, but the proximal portion of the embolic coil failed to break.¿the last part of the coil did not perform loops anymore".The coil was ¿supplied¿ with a 3mm non-cerenovus coil, which was considered too small.There was no report of consequence of impact to the patient.Multiple attempts were made without success to obtain additional information.The micrusframe was not returned for investigation.In addition, the sterile lot number was not available; therefore, a review of the manufacturing documentation could not be performed.Procedural imaging provided by the customer was reviewed and the results are as follows: films showed the coil being inserted into an aneurysm with catheter pushback.The catheter pushback prevented the coil from being deployed into the aneurysm.It would be expected for the coil to fit into the aneurysm with the proximal portion of the coil to fold/coil as it was pushed into the aneurysm.Instead, the microcatheter backed out as the last portion of the coil was being delivered.The size of the coil may be appropriate for the aneurysm, but this coil was too long to allow full deployment into the aneurysm.An alternative approach may have been to advance the microcatheter further into the aneurysm, or, the use a shorter coil than the 5.4 cm coil used.The push back of the microcatheter as the proximal end of the coil is inserted into the aneurysm is very common with all systems.While it may indicate lack of support by the microcatheter it is still commonly observed with many different types of microcatheters, firm or soft.There was no evidence of a device malfunction observed, and the event appears to be related to device selection.Based on the review of procedural films, the reported customer complaint of positioning difficulty-poor conformability was confirmed.Positioning difficulty is a known procedural complication associated with the micrusframe thermo-mechanical coil.The root cause of the event could not be conclusively determined; however, review of the available information suggests that clinical and procedural factors, including aneurysm size/vessel characteristics, device selection, device manipulation, and the concomitant microcatheter, may have contributed to the reported issue.There is no evidence of manufacturing or design issues that contributed to the event.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRUSFRAME10 3MM X 5.4CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
REFER TO SECTION H10
47709 fremont blvd
fremont CA 94538
MDR Report Key7789337
Report Number3008114965-2018-00713
Device Sequence Number1287172
Product Code KRD
UDI-Device Identifier10886704077800
UDI-Public10886704077800
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMFR100305
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/26/2018
Supplement Date Received by Manufacturer08/27/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3MM NON-CERENOVUS COIL
-
-