• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, LATERAL; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, LATERAL; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 0279350201
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.Mfg date: the device manufacture date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the tipbroke during a case.All pieces were retrieved and the procedure was completed successfully.
 
Event or Problem Description
It was reported that the tip broke during a case.All pieces were retrieved and the procedure was completed successfully.
 
Additional Manufacturer Narrative
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: during acl, the tip of the probe broken off.It was considered all the broken tips/flakes could be retrieved.The failure identified in the investigation is consistent with the complaint record.The probable root cause could be device came in contact with another hard instrument.Excessive force used when inserting or removing the probe, probe inserted into the obstructed passageway or any forceful impact to the tip of the probe with rigid objects.The probe used as a tool for mechanical displacement of tissue or bone, or to pry or scrub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES, LATERAL
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key7789469
Report Number0002936485-2018-00580
Device Sequence Number1535794
Product Code GEI
UDI-Device Identifier07613327056914
UDI-Public07613327056914
Combination Product (Y/N)N
PMA/510(K) Number
K071859
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0279350201
Device Lot NumberNR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2018
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer07/20/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date11/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-