| Model Number |
231235 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported the sheath peeled during insertion and removal of the fiber scope.There were no missing pieces.It was later reported on (b)(6) 2018, that the outside of the tube of the sheath may have peeled.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.The sample received was used and in good condition.The reported event was difficult to observe on the complaint sample.The sample underwent evaluation using 20x magnification, and it was observed that the outside of the sheath had a very minute flaking.Processes already exist during manufacturing and quality control to eliminate any defects.There is a 100% visual inspection during production.Quality control testing includes a functional testing of coefficient of friction and congo red test to identify any coating discrepancies.A review of the dhr analysis data indicated no issues for the referenced lot.The failure could not be replicated and a root cause could not be determined.The instructions for use were found adequate and state the following: "description: the proxis¿ ureteral access sheath is a two component ureteral dilation system which contains a single lumen for injection of fluids as well as passage of endoscopes and related instruments.The packaged product includes the following items: 1 - hydrophilic-coated dilator with female luer connector 1 - hydrophilic-coated sheath with hub indications for use: the proxis¿ ureteral access sheath is indicated for use in endoscopic urology procedures where ureteral dilation and ureteral access is desired for injection of fluids and insertion and removal of endoscopes and related instruments.Contraindications: ¿ patients who are contraindicated for retrograde urological procedures.¿ patients who are contraindicated for antegrade urologic procedures, including but not limited to patients with blood clotting anomalies due to coagulopathies or pharmacological anticogagulations.¿ patients who have the presence of tight strictures which would limit use of the device.¿ patients who have the presence of large obstructing distal ureteral calculi.Warning: for single use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/ or lead to device failure, which in turn, may result in patient injury, illness or death.Reuse reprocessing or re-sterilization may also create a risk of contamination of the device and/or cause patient infection or crossinfection, including, but not limited to, the transmission of infectious disease(s) from one patient to another.Contamination of the device may lead to injury, illness or death of the patient.Do not use if sterile barrier is damaged.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and with applicable laws and regulations.".
|
| |
|
Event or Problem Description
|
|
It was reported the sheath peeled during insertion and removal of the fiber scope.There were no missing pieces.It was later reported on 7/29/2018, that the outside of the tube of the sheath may have peeled.
|
| |
|