| Catalog Number |
PLC231000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
07/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to endoleak and aneurysm enlargement.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2015, the patient was treated for an aortic abdominal aneurysm with gore® excluder® aaa endoprostheses.On (b)(6) 2018, imaging identified a distal type i endoleak on the plc231000 with aneurysm enlargement of an unknown amount.It was reported the endoleak was caused by disease progression (dilatation) causing a loss of seal.On the same day, the right external iliac artery was coiled embolized and a device was implanted to extend from the plc231000 to treat the endoleak.Final imaging identified that the endoleak was resolved.The patient tolerated the procedure.
|
| |
|