| Model Number |
DVBC3D4 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patients right ventricular (rv) lead had triggered a lead integrity alert (lia) with increased sensing integrity counter (sic), high rate non-sustained episodes, noise, high thresholds, low defib impedance, and non-capture.It was thought that there could be a connection issue in which the lead had not been fully inserted into the implantable cardioverter defibrillator (icd) header or the lead is dislodged.A revision was completed and the device was extracted and replaced, while the rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
| |
|