Catalog Number 12017709122
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event occurred in: (b)(6).
 
Event or Problem Description
The customer complained of questionable thyroid results for 1 patient tested on a cobas 6000 e 601 module compared to competitor (abbott and ria) results.The customer stated that the patient's elecsys t3 and elecsys t4 assay results were similar before and after t3/t4 supplementation and they did not consider that reasonable.From the data provided, a reportable malfunction was provided for t3, t4, elecsys ft4 ii assay, and elecsys tsh assay.The specific date of testing for the data that is a reportable malfunction is not known, but clarification has been requested.This medwatch will cover the t4 data.Refer to medwatch with patient identifier (b)(6) for information on the t3 results.Refer to medwatch with patient identifier (b)(6) for information on the ft4 ii results.Refer to medwatch with patient identifier (b)(6) for information on the tsh results.The customer did not provide the thyroid data for the time period before the patient was taking t3/t4 supplements.The customer performed multiple dilutions of different dilution factors using two different diluents, but product labeling does not support dilution testing.The erroneous results were not released outside of the laboratory.There was no allegation of an adverse event.The cobas e601 serial number was (b)(4).The investigation is currently ongoing.
 
Additional Manufacturer Narrative
The patient sample was not available for further investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Additional Manufacturer Narrative
The specific date of testing for sample a was not provided, but it was mentioned that the testing on the customer's cobas e602, the abbott system, and the ria method were within three days of each other.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS T4 ASSAY
Common Device Name
RADIOIMMUNOASSAY, TOTAL THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7790435
Report Number1823260-2018-02742
Device Sequence Number11060008
Product Code CDX
UDI-Device Identifier04015630912728
UDI-Public04015630912728
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K961490
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number12017709122
Device Lot Number248077
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer07/30/2018
07/30/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date08/29/2018
10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
T3/T4 SUPPLEMENTS
Patient Age53 YR
-
-