TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Customer disconnected call.Unable to contact customer back.Most likely underlying root cause: mlc-20-user's test strip had poor storage(bathroom).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Unable to send product notification letter to customer to contact customer care.No address on file.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.Customer is concerned with tests results of 127 and 82 mg/dl.The customer's expected fasting blood glucose test result range is 150 -250 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the bathroom and in customer's purse.The test strip lot manufacturer's expiration date is 06/29/2019 and open vial date is two months.Customer had disconnected call.The meter memory was reviewed for previous test result history: result 1: 127mg/dl, date: (b)(6) 2018, time: 7:50am, fasting.Result 2: 82mg/dl, date: (b)(6) 2018, time: 7:27pm, fasting.Customer disconnected call with ccr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7790741
Report Number1000113657-2018-00868
Device Sequence Number16523223
Product Code NBW
UDI-Device Identifier00021292007911
UDI-Public(01)00021292007911
Combination Product (Y/N)N
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/29/2019
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2729
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2018
Initial Report FDA Received Date08/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-