| Model Number |
CIE3-13-24-00 |
| Medical Device Problem Codes |
Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Pain Relief (2388)
|
| Date of Event |
06/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Our investigation of this event with the information to date suggests that the event is due to potential inherent instability in the patient's spine that caused a slip at l4 and pressed (or caused the device to press) on the patient's nerve causing pain four months post-op.There is no evidence of device malfunction or direct device cause/contribution to the need for revision surgery.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision of a varilift-lx device was preformed due to intractable back pain in a patient 4 months post-op.
|
| |
|
Search Alerts/Recalls
|