No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that after replacing the surgeon monitor and liquid crystal display (lcd) the system was functioning as intended.The system then passed the system checkout and was found to be fully functional.The surgeon monitor and lcd were returned to the manufacturer for analysis.Analysis found that the returned monitor powered up with a normal display.After running for an extended time the display went blank as reported.The returned cable was found to be in good condition with no apparent physical damage and passed a continuity test with no opens or shorts detected.Analysis found that the reported event was related to a computer component issue.
|