• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM CAT3 ASPIRATION CATHETER; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM CAT3 ASPIRATION CATHETER; DXE Back to Search Results
Catalog Number CAT3
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the cat3 was fractured approximately 102.0 cm from the hub.Conclusions: evaluation of the returned cat3 revealed that the device was fractured.If the device is forcefully retracted out of its packaging hoop at an extreme angle, damage such as a kink may occur.Further manipulation of the kinked device during preparation, may result in a fracture.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital staff found that there was a kink in the indigo system cat 3 aspiration catheter (cat3) and it looked like it was coming apart upon removal from the packaging.The damage to the cat3 was found prior to use.Therefore, the cat3 was not used in the procedure.The procedure was completed using a new cat3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CAT3 ASPIRATION CATHETER
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7791838
Report Number3005168196-2018-01617
Device Sequence Number14037512
Product Code DXE
UDI-Device Identifier00814548016085
UDI-Public00814548016085
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/20/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2021
Device Catalogue NumberCAT3
Device Lot NumberF81542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2018
Initial Report FDA Received Date08/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-