• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL Back to Search Results
Model Number M006180155170
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a contour vl stent was opened for use for a procedure to be performed in the kidney on (b)(6) 2018.According to the complainant, during unpacking, it was noticed that the stent was broken into two pieces along the shaft of the stent.The procedure was completed with another contour vl stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Additional Manufacturer Narrative
Correction to model number from 64335 to m006180155170.Correction to mfr contact first name from (b)(6) to (b)(6) and added (b)(6) to mfr contact last name.Correction to device problem code.Problem code 1069 captures the reportable event of stent shaft break.
 
Event or Problem Description
It was reported to boston scientific corporation that a contour vl stent was opened for use for a procedure to be performed in the kidney on (b)(6) 2018.According to the complainant, during unpacking, it was noticed that the stent was broken into two pieces along the shaft of the stent.The procedure was completed with another contour vl stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7791847
Report Number3005099803-2018-60062
Device Sequence Number19420159
Product Code FAD
UDI-Device Identifier08714729765806
UDI-Public08714729765806
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K974541
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model NumberM006180155170
Device Catalogue Number64335
Device Lot Number0021945964
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/23/2018
Supplement Date Received by Manufacturer09/24/2018
Initial Report FDA Received Date08/16/2018
Supplement Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-