| Model Number |
ZCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Implant and explant dates: if implanted or explanted, give date: not applicable, as the lens was inserted and removed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the doctor was trying out a smaller blade to make incision enlargements before cataract implantation.The doctor inserted the insertion system, encompassing a zcb00 28.0 diopter lens, into the ''smaller'' incision site and as the lens advanced into the patient's right eye (od), the leading haptic caught on the incision site.Therefore, the insertion system/lens was removed, the incision site was enlarged, and then replaced with a back up lens of the same model and diopter.Reportedly, there was no patient injury and the patient is doing fine post-operatively.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
In review, the initial emdr was reported as reportable incorrectly.The event occurred due to a handling problem where the physician used a smaller blade to purposely make the incision size smaller, which is not reportable.Therefore, no more reporting will be filed under mdr report number 2648035-2018-01111.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-(b)(4) and capa-(b)(4).
|
| |
|