MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Myocardial Infarction (1969); Thrombus (2101); No Known Impact Or Consequence To Patient (2692)
Date of Event 09/27/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
It was identified that during a coronary atherectomy procedure using a csi orbital atherectomy device (oad), a thrombus occurred.This was identified during an optical coherence tomography (oct) image review following the procedure.The procedure was completed with no adverse events related to thrombus reported.
 
Event or Problem Description
The observed thrombus was located at stent.No thrombus was observed on angiography.
 
Additional Manufacturer Narrative
Per physician review, this observation of thrombus was not associated with any adverse event to the patient and was a subclinical finding associated with the intravascular imaging core lab analysis required as part of a clinical trial.Based on this review, this event no longer meets the definition of a reportable serious injury.Csi id (b)(4).
 
Additional Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for use manual states that myocardial infarction is a potential adverse event that may occur and/or require intervention with use of the system.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Event or Problem Description
Additional information received indicated the clinical event committee reviewed images and concluded that the use of the diamondback 360 coronary orbital atherectomy device may have contributed to a myocardial infarction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Common Device Name
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
Manufacturer Contact
laramie otto
1225 old highway 8 nw
saint paul, MN 55112
MDR Report Key7793064
Report Number3004742232-2018-00252
Device Sequence Number1291114
Product Code MCX
UDI-Device Identifier10852528005169
UDI-Public(01)10852528005169(17)190831(10)199867
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P130005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Model NumberDBEC-125
Device Catalogue Number70058-02
Device Lot Number199867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer03/01/2019
12/04/2023
Initial Report FDA Received Date08/17/2018
Supplement Report FDA Received Date03/14/2019
01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age74 YR
Patient SexUnknown
-
-