• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD CHROMIC GUT; SUTURE, ABSORBABLE, NATURAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD CHROMIC GUT; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Model Number G-2797K
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during circumcision, on closure of the fore skin, the needles detached from the suture as the surgeon was suturing.A new device was opened and used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) received thirty-one devices.A tactile and microscopic inspection of the sutures was performed with no abnormalities.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMIC GUT
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7793113
Report Number9612501-2018-01488
Device Sequence Number16537494
Product Code GAL
UDI-Device Identifier10884521054707
UDI-Public10884521054707
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K885018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberG-2797K
Device Catalogue NumberG-2797K
Device Lot NumberD7H1524ZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Initial Date Received by Manufacturer 08/02/2018
Supplement Date Received by Manufacturer01/24/2019
Initial Report FDA Received Date08/17/2018
Supplement Report FDA Received Date02/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-