U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC 6F RADIOLOGY BASIC TRAY (DUAL-LUMEN) (135 CM GUIDEWIRE); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC 6F RADIOLOGY BASIC TRAY (DUAL-LUMEN) (135 CM GUIDEWIRE); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number 3276135
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of recn1189 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the as the nurse and doctor were prepping the picc for insertion, it was found the catheter hub clamp was faulty and leaking.A new set was opened and the procedure was completed without further incidence.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged clamp was confirmed and the cause appeared to be supplier related.One 6 fr d/l powerpicc was returned for investigation.The catheter had been trimmed at the 43cm depth mark.The clamp on the extension leg with the purple connector was damaged.A microscopic examination of the clamp revealed material deformation that was consistent with shearing.It appeared that the clamp was damaged during the molding process.No damage was observed on the adjacent clamp tag.A functional test revealed that fluid bypassed the closed clamp.A supplier incident report was sent to supplier, regarding this incident, as the clamp is a supplied component.A lot history review (lhr) of recn1189 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the as the nurse and doctor were prepping the picc for insertion, it was found the catheter hub clamp was faulty and leaking.A new set was opened and the procedure was completed without further incidence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC 6F RADIOLOGY BASIC TRAY (DUAL-LUMEN) (135 CM GUIDEWIRE)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7793312
Report Number3006260740-2018-02114
Device Sequence Number11281708
Product Code LJS
UDI-Device Identifier00801741028007
UDI-Public(01)00801741028007
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K050931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number3276135
Device Catalogue Number3276135
Device Lot NumberRECN1189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Event Location Hospital
Initial Date Received by Manufacturer 08/08/2018
Supplement Date Received by Manufacturer08/24/2018
Initial Report FDA Received Date08/17/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-