(b)(4) if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was discarded by the customer therefore the device is not available for a physical evaluation.This complaint is not confirmed.It was reported that the drill bit was dull and could not go through the bone during use.No further information regarding the procedure or the device used has been provided to determine a root cause for this failure.Furthermore, no lot number was supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.Based on the information available this complaint will be accounted for and monitored via post market surveillance activities.However, if additional information is made available, the investigation will be updated as applicable.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).The lot number is currently unavailable.Associated medwatch: [mr#1221934-2018-54510].
|