DEPUY MITEK VAPR ARCTIC SUCTION ELECTRODE WITH DOME TIP; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE USE
|
Back to Search Results |
|
| Catalog Number |
225029 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/08/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, therefore unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the affiliate in (b)(6) that during arthroscopic hip scope procedure, the surgeon noticed that the electrode was not responsive about 30 minutes into the procedure, according to the report, the electrode's tip articulated and was not working properly.It was further reported that when viewed with the c-arm, it was clear that the tip was broken.It was reported that the surgeon opened another arctic electrode to finish the case.It was reported that the patient was fine post-surgery.It was reported that there was no delay in the surgical procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|