An event regarding crack/fracture of the ps post involving a scorpio insert was reported.The event was confirmed based on evaluation of the returned device.A material analysis was performed on the returned device and concluded that the insert fractured at the post.The fracture initiated in overload, propagated in fatigue, and the final fracture occurred in overload.Burnishing, scratching and third-body indentations were observed on the insert condyles.These are common damage modes of uhmwpe.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: not performed as medical records were not provided for review.All devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the reported lot.A material analysis was performed on the returned device and concluded that the insert fractured at the post.The fracture initiated in overload, propagated in fatigue, and the final fracture occurred in overload.Burnishing, scratching and third-body indentations were observed on the insert condyles.These are common damage modes of uhmwpe.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Further information such as patient details, operative notes, orthopedic clinic notes and x-rays would be required to determine the root cause for this event.If additional information becomes available, this investigation will be reopened.
|