U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 740 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 740 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 740
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was turned on and the power supply was smoked.The service engineer installed a new power supply.The ventilator passed all testing per manufacturing specification and was placed back into clinical use.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the 740 ventilator after turning on the device "power supply smoked".The ventilator was not in use on a patient at the time of the reported event. .
 
Additional Manufacturer Narrative
Device evaluation summary: the replaced power supply was returned to medtronic for analysis.It was observed that there was thermal damage on the printed circuit board (pcb), which is the likely cause of the reported condition.The power supply that was returned is now obsolete and no schematics, design files or replacement parts were available so no further investigation was possible.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
740 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway,gw
MDR Report Key7794147
Report Number8020893-2018-00376
Device Sequence Number1143903
Product Code CBK
UDI-Device Identifier10884521078369
UDI-Public10884521078369
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K990897
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number740
Device Catalogue NumberG-061792-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Initial Date Received by Manufacturer 07/27/2018
Supplement Date Received by Manufacturer10/12/2018
Initial Report FDA Received Date08/17/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-