Catalog Number 359.217
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information is not available for reporting.Date of event is unknown.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that the cutter broke.It is unknown if there was surgery or patient involvement.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Part: 359.217 , lot: l042551, manufacturing site: hägendorf, release to warehouse date: 25.Aug.2016.The device history record shows this lot of (b)(4) pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Only top level of the device history record reviewed as sub-components are not lot tracked review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition customer quality reviewed the photograph provided.Product was not returned.The received pictures are in a poor quality which makes it difficult to make a clear statement.There are two parts visible.By this i can only assume that a breakage occurred and therefore the complaint will be rated as confirmed.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTER FOR TI ELASTIC NAILS
Common Device Name
INSTRUMENT,CUTTING,ORTHOPAEDIC
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7794837
Report Number8030965-2018-55804
Device Sequence Number1293371
Product Code FZT
UDI-Device Identifier07611819096646
UDI-Public(01)07611819096646
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number359.217
Device Lot NumberL042551
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date08/17/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-