The device was in use for treatment.The device was not returned for analysis, therefore, the clinical observation could not be confirmed.Based on the investigation, a capa is not required.A change request was initiated to implement action to prevent recurrence of this event.Oscor will continue to monitor this event type.The event will be re-evaluated if additional information becomes available.Per qa procedure adelante-s introducer sheath in process and final inspection: 13.3.2 break and peel test: using samples from fit check as described in 13.3.1, manually break the sheath and hub and verify that the seal splits easily without extreme elongation.Also verify that the split cap and seal remain secure and do not break free or loosen from the sheath hub.Manually peel the sheath and verify the sheath peels easily along the sheath body and tip.Per ifu : sharply snap the tabs of valve housing in a plane perpendicular to the long axis of the sheath to split the valve and peel sheath apart while withdrawing from the vessel.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|