| Catalog Number |
1125250-12 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Angina (1710); Thrombosis (2100)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.It should be noted that the reported patient effect(s) of angina and thrombosis are listed in the xience alpine, everolimus eluting coronary stent systems instructions for use (ifu) as a known patient effect(s) of coronary stenting procedures.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.The other two xience alpine stents are filed under separate manufacturing report numbers.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, two 2.5x12mm xience alpine stents were implanted, one in the 2nd obtuse marginal and one in the 3rd obtuse marginal coronary arteries.A 2.5x23mm xience alpine stent was also implanted in the 1st diagonal coronary artery.Reportedly, stent placement angiographic appearances were acceptable.Post procedure and prior to hospital discharge, the patient reported abdominal cramping and tridil infusion was provided.Chest pain grade 9/10 was reported and aggrastat infusion was initiat ed.Another angiogram was performed.Stent thrombosis was noted in the proximal left anterior descending artery, 1st diagonal, mid circumflex, and ostial 2nd obtuse marginal.Another percutaneous intervention was performed as treatment with additional stent placement.The event resolved without sequela.There was no device malfunction.No additional information was provided regarding this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Correction: report source - customer representative unchecked.
|
| |
|
Search Alerts/Recalls
|