| Model Number |
A28-28/C95V |
| Medical Device Problem Codes |
Leak/Splash (1354); Migration or Expulsion of Device (1395); Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in the event will not be returned for evaluation as it remains implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the patient was implanted with a bifurcated stent graft and an infrarenal stent graft extension to treat an abdominal aortic aneurysm.On (b)(6) 2018, endologix was made aware of a type 1a endoleak with stent migration.On (b)(6) 2018, the physician elected to implant a suprarenal extension to resolve this leak.The patient was reported as doing fine and there have been no additional patient sequelae reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain relevant medical records and images.As such, event determination, off label conditions, related patient harms and patient disposition could not be independently assessed.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
|
| |
|