| Model Number |
MMT-523NAL |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Hypoglycemia (1912)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call on (b)(6) 2018 that the infusion set had a bent cannula.Customer's blood glucose at the time of the incident was 48mg/dl.Product is not expected to return.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received unable to perform all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test due to completely broken reservoir tube lip.(b)(4).
|
| |
|