| Catalog Number |
120095 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The cement tool bent intraop.Case type: pka.
|
| |
|
Event or Problem Description
|
|
The cement tool bent intraop.Case type: pka.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding damage involving a mako instrument was reported.The event was not confirmed.Visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.No medical records were received for review with a clinical consultant.Review of the product history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for this lot.It was reported that the cement removal tool bent during surgery.The event could not be confirmed, nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.No further investigation for this event is possible at this time as no devices and/or insufficient information was received by stryker orthopaedics.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|